A modified CO2/O2 Guedel airway improves capnographic accuracy compared with a CO2/O2 nasal cannula: An infant manikin study
European Journal of Anaesthesiology Jul 12, 2018
Moll J, et al. - Researchers used a manikin model of a 6-month-old infant to compare capnography from a modified Guedel airway with a CO2 port located at the tip with that from a CO2/O2 nasal cannula. They focused on the comparison of differences between tracheal and device CO2 as well as on the impact of various respiratory settings and O2 flows on the CO2 difference. They found that accuracy of capnography traces derived from a sample port at the tip of a modified CO2/O2 Guedel airway was better than those obtained from a CO2/O2 nasal cannula.